Frequently Asked Questions about Cogent TYPE-C HPLC Columns
- What is the biggest difference between current HPLC phases and TYPE-C columns?
- How do I store TYPE-C columns?
- How do I use these columns in Normal Phase after I have used them in Reverse Phase?
- How do I condition TYPE-C Columns for use?
- What is Silica-Hydride?
- What Aqueous Normal Phase Chromatography?
- Many other column suppliers suggest different columns for Acids, Neutrals or Bases, which TYPE-C column is best for each of these compound groups?
- Cholesterol is a large molecule, is it stable as a stationary phase?
- Why should I use a TYPE-C column to develop methods when I already have experience with Type-B columns?
- Is TYPE-C Silica high purity and low metal content silica?
- Will TYPE-C Silica have the same physical properties as my HPLC column does now?
- What are the advantages of Silicon-Carbon bonding for stationary phases? How will that benefit me?
- What is on-column hydrolysis and what causes it?
- Is TYPE-C Silica a hybrid particle?
- If I want to do Normal Phase with these columns, do I still use Phosphoric Acid in the solvents?
- How do I use an acid as an additive if I scale up to prepartive work?
- I used a TYPE-C HPLC Column and dissolved my samples in Acetonitrile and water. Can I use other solvents such as methanol, IPA or Ethyl Acetate?
- I am working with LC/MS, can I substitute Formic Acid for Phosphoric Acid with Cogent TYPE-C columns?
- Which is better with TYPE-C Silica based HPLC Columns: TFA, Phosphoric Acid or Formic Acid?
- What is the "Batch Reproducibility" of the Cogent Bidentate C18?
- Since you manufacture the Cogent Type-C silica using a transition metal catalyst, what if any is the presence of any trace quantity on the surface and how did you measure this?
- What are the main differences between HILIC columns and a Cogent TYPE-C silica based column?
- What are the concentrations of the test analytes in the actual test chromatogram that comes with the column?
- Can I use Isopropyl alcohol as solvent in the mobile phase?
- I am working with plasma extracts and we are diligent to remove our guard columns regularly but we would still like to perform a rigorous column washing to minimize contamination and carry over from run to run. What can you suggest other than extra column volumes of the mobile phase?
- What other types of solvents can you suggest for rigorous column washing to remove adsorbed materials?
- Most acids and bases are not effective when the organic content of the mobile phase goes too high. Since I am interested to use the Cogent columns in ANP, what can I do to adequately "acidify" the mobile phase when organic is higher than 40%?
- I was trying to reproduce the separation of amino acids from an application note on your website and the retention times for the individual compounds were different. How do I get the same results as in your application note?
- I injected 100ul of my sample onto a DH column. The resulted peaks of amino acids were very distorted. How can I improve this?
- I was using a Cogent Diamond Hydride column and I prepared solvents for positive ionization LCMS. Solvent A:50% methanol/50% DI water + 1% formic acid and solvent B:90% acetonitrile/10% DI water/0.1% formic acid. I started the gradient from 100% A and over 10 minutes went to 50%. Polar metabolites were not retained. What did I do wrong?
- I was using a Cogent Diamond Hydride column with LCMS and solvents containing 10mM ammonium acetate. I noticed that my source was getting a white film on it. At the same time the background noise was very low. Is my column losing the packing material?
- I was analyzing maleic and fumaric acids in biological extract using Cogent DH column. After 10 perfect injections (%RSD of retention times around 0.5), I noticed that the peaks intensity diminished about 10 times, however the peaks still have the same retention time. What happened to the column?
- A Cogent Diamond Hydride column was used for analyzing AMP, ADP, ATP, UDP and GPT in biological extracts. After 2 days of perfect chromatograms my peaks started to deteriorate. What can I do to get back initial results?
1. What is the biggest difference between current HPLC phases and TYPE-C columns? [top]
Over 95% of the surface silanols of our high purity silica have been replaced with silicon-hydride (Si-H)
producing a stationary phase that adsorbs and de-sorbs solvents very differently than your current silica HPLC
columns. This also means that the surface of the TYPE-C silica is slightly hydrophobic and therefore does not have a strong relationship
with water and thus there is no "hydration shell" or water layer on the surface as with all ordinary silica. This feature alone
is a compelling reason to use this material for the many benefits you can gain in HPLC.
Some other differences in our TYPE-C HPLC columns and ordinary silica based columns: retention of polar compounds that normally
will not retain; retaining polar and non polar compounds in the same isocratic run; performing very rapid gradients with very fast
equlibration; performing “inverse gradients”; performing organic normal phase with bonded phases such as C18 without drying out
solvents; excellent peak shapes without end-capping and incedibly long column life that will save you money in the cost of
acuisition as well as the cost of use and ownership of the columns. For more information on the differences, please
click here.
2. How do I store TYPE-C columns? [top]
The TYPE-C™ phases can be stored in aqueous / organic solvents. For example: 90% methanol/10% DI water, or 90% acetonitrile/10% DI water. Columns with TYPE-C™ phases can also be stored in organic solvents (methanol, acetonitrile).
Long Term Storage:
- 1.Rinse the column with 8 column volumes with pure organic solvent that you use as your mobile phase. It is important to use a non-buffered mobile phase.
- 2. Cap both ends of the column with the provided protective end caps or with Super-Link (Seal-Links) End Fittings and capping them with column plugs provided.
- 3. Return column to the box it came in for safekeeping.
- 4. Store column at room temperature.
- 5. Never let the column dry out.
- 6. Never store the column in buffers.
- 7. Never store the column in a liquid that is more than 30% water.
Overnight:
- 1. Rinse the column with 8 column volumes with pure organic solvent that you use as your mobile phase. It is important to use a non-buffered mobile phase.
- 2. Cap both ends of the column with the provided protective end caps or with Super-Link End Fittings and capping them with the column plugs provided.
3. How do I use these columns in Normal Phase after I have used them in Reverse? [top]
Just as a reminder, all Cogent Bidendate C18 and C8 columns are filled with water containing solvents during shipping. To start work in Normal Phase Mode, a simple switching procedure is recommended.
A simple 30 minute procedure allows switching from one mode to another.
Procedure:
A – moving from Reverse Phase to Normal Phase HPLC; pump 100% methanol for 15 minutes at 1 mL/min. flow rate, followed by 15 minutes 100% methylene chloride. The column is ready to be equilibrated with the mobile phase for NP-HPLC.
B – moving from Normal Phase to Reverse Phase HPLC; pump 100% methylene chloride for 15 minutes at 1 mL/min. flow rate, followed by 15 minutes 100% methanol. The column is ready to be equilibrated with mobile phase for RP-HPLC.
4. How do I condition TYPE-C Columns for use? [top]
The best and most efficient way to condition the TYPE-C™ HPLC Columns is to run approximately 6 bed volumes of the mobile phase you are going to use in your run through the column. It is very important to note that if the column was stored with solvents that might be immiscible with this mobile phase, you must first flow a commonly miscible solvent, such as isopropanol, through the column for 15 bed volumes, prior to re-equilibration in the aqueous or organic based solvent to be used next. Be careful that the back pressure does not become excessive when purging with IPA as this is a more viscous solvent which can generate high back pressures even at low flow rates.
Then you should inject either a known standard or one of the standards used in the actual test chromatogram supplied with the column with appropriate mobile phase or, in your mobile phase and re-run until you get reproducible and acceptable results. If you use one of the standards and the test chromatogram and you cannot reproduce k’ or N/meter, contact us before proceeding and we will advise you on possible remedies.
If you are working with a sample for the first time, it is suggested that you first equilibrate as described above with a compound that is known to you to establish equilibration. Then you should run your unknown at least two times.
5. What is a Silica-Hydride? [top]
This is our term for the surface of TYPE-C Silica which is dominated by silicon-hydride groups (Si-H). We are
referring to the Silica Surface and the Hydrides.
6. What is Aqueous Normal Phase (ANP)? [top]
Aqueous Normal Phase (ANP) is a full, new mode of HPLC. Reverse Phase (RP) and Normal Phase (NP) HPLC are more well
known but ANP offers chromatographers benefits that are not possible with other modes.
In Aqueous-Normal Phase, the maximum retention time of target compounds is with 100% acetonitrile (least polar
solvent) and as you increase the polar solvent content (Aqueous), the retention reduces to a minimum when the mobile
phase is between 50 - 20% Acetonitrile (depending on the type of polar compound). You get the Normal Phase mode using Aqueous or Reverse Phase Solvents. Polar compounds
that would not retain in RP for example, retain very well in ANP.
By definition:
Normal-Phase Chromatography (NP)is a chromatographic mode that will increase the retention of a target compound as
the mobile phase becomes less polar (a decrease in concentration of the most polar solvent, often water) conversely
the compound has longest retention in a non-polar solvent such as 100% hexane. The main separation mechanism is based
on the functionality of the analytes.
Reverse-Phase Chromatography (RP) is the opposite or the reverse of the above; there is an increase in retention time
of the target compounds as the mobile phase becomes more polar (increase concentration of the most polar solvent, often
water). The main separation mechanisms of RP is based on the hydrophobicity of the analytes.
Aqueous Normal-Phase Chromatography (ANP) is defined as a normal phase separation pattern using the reverse phase
solvents Water and Acetonitrile. The main separation mechanisms of ANP is based on the functionality of the analytes.
7. Many other column suppliers suggest different columns for Acids, Neutrals or Bases, which TYPE-C column is best for each of these compound groups? [top]
The TYPE-C HPLC phases are unique in that they have different retention mechanisms which can be exploited to
give unique separations. It may not be necessary to use different columns for different classes of compounds.
However it is recommended to use two Diamond Hydride columns for separation of polar metabolites in biological fluids., one for use with mobile phases containing formic or acetic acid and second column for mobile phases containing ammonium acetate or formate.
8. Cholesterol is a large molecule, is it stable as a stationary phase? [top]
Yes, the UDC Cholesterol phase has proven to be stable in 0.1% v/v TFA and 0.1% v/v H3PO4 at
acid pH’s and up through pH 8.00 with 20mM NH4COOH buffer with NH4OH.
Note: When using TYPE-C™ HPLC columns often it is not necessary to use high pH (whch is harmful to HPLC instruments). Many challenging separations can be achieved without using extreme pH.
A pH in excess of pH 8.00 should be avoided as this phase has a mono dentate (single point) attachment to the
silica matrix and an intermediate ester link in its structure. For higher pH’s the Cogent Bidentate C18 is
recommended.
9. Why should I use a TYPE-C column to develop methods when I already have experience with Type-B columns? [top]
Change is often work but can be worth every effort. With TYPE-C based columns, you can get good reverse phase
from them but within the same solvent system (aqueous) you can achieve normal phase retention and elution patterns
for polar compounds. You can also get polar and non polar compounds to retain and elute in the same isocratic run.
If gradients are required to cope with large differences in polarity of multiple compounds in a mixture the time it
takes to equilibrate the TYPE-C columns is almost negligible. This makes your method development very fast. Only TYPE-C
columns have Selectivity3 which means on the same column, you can do Reverse Phase, Normal Phase and Aqueous
Normal Phase.
Expand your skill set with TYPE-C based columns.
10. Is TYPE-C Silica high purity and low metal content silica? [top]
Yes, it is and it has all the advantages of high purity, type-b silica supports.
11. Will TYPE-C Silica have the same physical properties as my HPLC column does now? [top]
Yes, it will have all the mechanical strength, and durability that your modern, type-b packed HPLC column will have
but will exhibit other very beneficial properties such as pH and temperature stability that is superior to your
column.
12. What are the advantages of Silicon-Carbon bonding for stationary phases? How will that benefit me? [top]
Typical HPLC phases today are made with siloxane (Si-O-Si-C) bonds which are prone to hydrolysis. Silicon-Carbon
bonds are much more stable and resistant to hydrolysis and is therefore less phase bleed-off. This means that these
columns will last longer and be more rugged and will be reproducible longer. This also means that you might be able
to explore new mobile phase/additive and temperature conditions to achieve very difficult separations that might be
impossible with your current HPLC columns.
Also, standard type B columns made with silanization of organo-silanes to produce siloxane bonded ligands limits
the phases which may be bonded onto silica. With our patented, silica-hydride surface many compounds previous
thought to be improbable to bond to silica can be easily bonded. Examples of this is our unique
UDC-Cholesterol, plus our unique ability to custom synthesis novel phases for our customers.
With TYPE-C HPLC phases you will pay much less per sample(injection) than with any other HPLC column.
13. What is on-column hydrolysis and what causes it? What will be the effect on TYPE-C Silica phases? [top]
Hydrolysis is a common degradation mechanism, which is facilitated by the presence of acids or bases in solution or
on the surface of solids the liquid is in contact with. With solids in contact with liquids, the larger the surface
area the greater the potential for hydrolysis. Irregular, type A & type-B HPLC phases all have by design very high
surface to volume ratios plus Si-OH groups populate this surface. Even a fully end capped traditional phase can
have 30 to 50% free silanols on its surface. Using photodiode array detectors and HPLC/Mass Spec has highlighted
this degradation activity.
Certain analytical disciplines such as Natural Product Bio-actives and Forensic Science have shown that 10 to 20%
of bio-active compounds may hydrolyze quantitatively “on-column” due to the conditions inside the HPLC Column.
Historic analyses were usually made by these disciplines, of degradation products and not the injected material due
to this activity in the column. TYPE-C Silica based products with silica-hydride surface (Si-H) are much less
hydrolytically active than the historic silanol (Si-OH) populated phases and may not hydrolyze certain compounds
making it possible to detect and track injected quantities instead of the degradation products inside created in
the column.
14. Is TYPE-C Silica a hybrid particle? [top]
No, it is an evolution of standard, high purity, based deactivated, type-B silica and if you are using a
commercially available silica based HPLC column, TYPE-C columns can be considered the same as what you are
currently using.
15. If I want to do Normal Phase with these columns, do I still use Phosphoric Acid in the solvents? [top]
Usually it would not be necessary to use pH selectivity (the reason for adding phosphoric acid) with organic normal
phase solvents, but this option is always option open to you for selectivity reasons. If chosen, it would be
typical to use an organic acid or base rather than inorganic additives.
16. Do I use an acid if I scale up to preparative work? [top]
With preparative, reverse phase applications with TYPE-C columns, it is always preferable to use un-buffered
eluents to reduce complications on evaporation. With TYPE-C columns significant volumes of non buffered
aqueous/organic will cause a moderate but acceptable loss of efficiency. It is therefore recommended to use weak
volatile buffers such as acetic acid/ammonium acetate/ammonium hydroxide mixes or their formate equivalents as
these are easily removed during evaporation. If this proves to be a problem for your method, un-buffered solvents
may be used, but the moderate loss of column efficiency will need to be allowed for and expected.
In most preparative applications reverse phase is rarely the first choice, as it is much easier to evaporate
organic solvents than water-based solvents. Organic normal phase (ONP) is usually the best choice to start with.
TYPE-C columns which can work in aqueous normal phase (ANP) also reduces the water content more easily on
evaporation, is therefore also a good choice for preparative work. In ONP it is rarely required to buffer the
eluent and the TYPE-C phases is stable to all un-buffered organic HPLC solvents tested to date.
17. I used a TYPE-C HPLC Column and dissolved my samples in Acetonitrile and water. Can I use
other solvents such as methanol, IPA or Ethyl Acetate? [top]
We do not advise TYPE-C users to use IPA for an eluent or as a solvent for target compounds. Regarding other
solvents, it is always best to try to dissolve your target in as close as possible to your isocratic eluent or the
start of your chosen gradient. If this is not possible gradually increase the percentage of the solvent in your
eluent that the target analyte is most soluble in.
18. I am working with LC/MS, can I substitute Formic Acid for Phosphoric Acid with Cogent
TYPE-C columns? [top]
Yes, DO NOT USE PHOSPHORIC ACID WHEN WORKING WITH LC-MS.
Use 0.1% Formic, 0.2% acetic or 0.005% TFA (Trifluoroacteic acid) acids.
19. Which is better with TYPE-C Silica based HPLC Columns: TFA, Phosphoric Acid or
Formic Acid? [top]
TFA will usually provide better peak shapes than Phosphoric or Formic Acid, but Phosphoric is normally recommended for UV detector use since it is less hazardous than TFA.
When using LC-MS, TFA can cause peak suppression, so it is not recommended as an additive to the solvents. Sometimes a small amount of TFA (0.05%) is added to the sample or to the mobile phase (0.005%). 0.1% Formic and 0.2% acetic acids can be used for positive ionization and 10 mM ammonium acetate (or ammonium formate) for negative ionization.
Please remember that ammonium acetate or ammonium formate are not soluble in organic solvents, so when you need to add these salts to acetonitirle follow the procedure outlined below:
A.Prepare a 200 mM solution of ammonium acetate (or ammonium formate) in DI water. Store in refrigerator.
B.Prepare solvent B (for ANP analysis) : 90% acetonitrile/10% DI water/10 mM ammonium acetate (or ammonium formate). For example to prepare 500 mL of solvent B mix together: 25 mL of 200 mM solution of ammonium acetate + 25 mL of DI water+ 450 mL of acetonitrile. When you follow this method of preparation of the mobile phase you will be able to achieve 0.5 %RSD when analyzing biological samples.
20. What is the "Batch Reproducibility" of the Cogent Bidentate C18? [top]
For 8 consecutive batches of Bidentate C18, carbon loading (excellent indicator for HPLC), did not vary more than
0.89%. This is almost within the experimental error of carbon load testing. Click here for more
information.
21. Since you manufacture the Cogent Type-C silica using a transition metal catalyst, what if any is the presence of any trace quantity on the surface and how did you measure this? [top]
This is an issue we confronted a long time ago. We were also concerned about this potential problem. First we
intentionally ran the bonding reaction at high catalyst concentration in order to actually reduce some of the Pt and
deposit it on the surface. This produced a visual confirmation, the surface became gray. Next we ran some compounds that
we knew would interact strongly with Pt and were able to observe longer retention and most important very distinct peak
tailing. Finally we determined the %Pt on the surface by ESCA spectroscopy and it was close to 1 atom %. Then we went ahead
and did our normal process. Of course the surface is very white under those circumstances. The compounds gave very good
retention and peak symmetry was virtually 1.0. Then we took the material and analyzed it by ESCA. There was no Pt detected
on the surface. The limit of detection is about 0.01 atom %. So we feel very confident that there is virtually no Pt
present as determined by both ESCA and chromatographic testing.
22. What are the main differences between HILIC columns and a Cogent TYPE-C silica based column? [top]
There are many differences between HILIC columns and Cogent TYPE-C columns but for the sake of brevity, I will
only answer with the main differences. The Cogent TYPE-C columns all perform similarly to HILIC as far as polar
compound elution order is concerned when using higher than 70% organic composition of the mobile phase. Both columns
perform separations that are based on variations of Normal phase called “Aqueous Normal Phase”. The HILIC stationary
phase is typically more polar than the TYPE-C bonded phase which is relatively non polar. HILIC columns will retain
polar compounds that cannot be retained by other columns and so will the TYPE-C columns. The HILIC columns will not
retain non-polar compounds. This is one of the big advantages for the TYPE-C columns, where you will be able to
separate polar and non-polar compounds in the same isocratic run.
On HILIC columns, polar compounds partition into and out of the hydration shell created by adsorbed water on the
silica surface. As the ACN concentration increases the water layer decreases and the charged polar analytes are
retained by the combination of cation exchange with the silanols under the water layer and the partitioning effect.
The combination of these two mechanisms retains polar compounds in HILIC mode.
On Cogent TYPE-C Columns, the charged polar compounds elute in a similar order as on HILIC columns as stated above.
However, non-polar compounds will be retained at the same time by the non polar ligand of the Cogent TYPE-C columns
which is the bonded phase. Since there are virtually no silanols to speak of on these columns, the polar compounds
are retained more by the adsorptive character of the silica-hydride and silica backbone which are now more available
to the analytes due to a very shallow solvent layer which is the organic component such as Acetonitrile, Methanol
etc. when at higher concentrations of the organic component.. Because of the lack of a “hydration shell” at high
organic content due the weak association with water of the TYPE-C silica, the shell will equilibrate and change more
rapidly with TYPE-C columns then with HILIC columns. This is an advantage for rapid gradients.
HILIC columns can only perform HILIC which is Aqueous Normal Phase while TYPE-C Silica based columns can perform
Aqueous Normal Phase, Reverse Phase and traditional Normal Phase using completely non-polar solvents such as Hexane.
There is no hysteresis when changing between these modes making this column extremely efficient.
To summarize:
Cogent TYPE-C columns for ANP chromatography outperform HILIC columns in the following areas:
•Retain nonpolar compounds by reversed phase mechanism
•Retain polar compounds by normal phase mechanism
•Both reversed phase and normal phase mechanisms can operate simultaneously
•Can separate samples with both polar and nonpolar compounds
•Equlibration time for gradient elution is 5 minutes between runs
•Repeatability for biological samples is around 0.5% RSD
23. What are the concentrations of the test analytes in the actual test chromatogram that comes with the column? [top]
It depends on the column and for the sake of brevity let me state the rationale we use to make up our standard test
mixture for the Cogent TYPE-C columns.
Starting with stock solutions and toluene, we make up a 10ml final working solution from which we inject. The formula
for making the final working solutions is as follows:
- Uracil Stock Solution
- 1mg/ml dissolved in mobile phase
- 100ul into 10ml of the final working solution (mobile phase)
- Anisole Stock Solution
- 200ul of Anisole dissolved in 20ml of mobile phase
- 800ul into 10ml of the final working solution
- Toluene-Straight
- 30ul of Toluene into 10ml of the final working solution
This is the starting point and depending on the column size, the detector and other non controllable parameters, a
further dilution of 2x up to 10x may be needed into the mobile phase.
24. Can I use Isopropyl alcohol as solvent in the mobile phase? [top]
We do not suggest using IPA (isopropyl alcohol) as component to your mobile due to the high back pressure you get from the high
viscosity. However, it will not harm the column and many people use IPA as a driving solvent when switching from RP to NP and
back and forth since it is miscible with both types of solvents. However, you should be extra careful to remove all of it from
your column before using it for analysis.
25. I am working with plasma extracts and we are diligent to remove our guard columns regularly but we would still like to perform a rigorous column washing to minimize contamination and carry over from run to run. What can you suggest other than extra column volumes of the mobile phase. [top]
When working with plasma extracts, we recommend using 50% methanol/50% DI water for washing between injections. Flow rates of 0.5ml/min
for 20-30 minutes for a 4.6mm x 250mm column should be sufficient.
When working with LC-MS and biological samples we recommend using 50% methanol/50% Di water as the A solvent. It helps clean the column between runs when biological samples are used.
26. What other types of solvents can you suggest for rigorous column washing to remove adsorbed materials? [top]
The Cogent TYPE-C columns are very stable and can be washed between runs with 100% Acetonitrile, DMSO THF and DMF among many
others. We use these solvents when typical washing procedures are not adequate. The only precaution I would give you is the
stay between pH 2 and pH 10 when using acids and bases.
27. Most acids and bases are not effective when the organic content of the mobile phase goes too high. Since I am interested to use the Cogent columns in ANP, what can I do to adequately "acidify" the mobile phase when organic is higher than 40%? [top]
For most of ANP separation of polar compounds it is enough to use 0.1% formic acid in both solvent A and B. If the peak shape is not satisfactory the addition of 0.05% of TFA to the sample is recommended. The other suggestion would be to use steep gradient.
28. I was trying to to reproduce the separation of amino acids form an application note from your web site and the retention times for the individual compounds were different. How do I get the same results as in application note? [top]
To obtain the same retention times for analyzed compounds as in application note gradient conditions have to be adjusted according to the instrument used. Each instrument has different plumbing (dimensions of the tubing), so obtaining exactly the same retention times is virtually impossible.
29. I injected 100 uL of my sample onto DH column. The resulted peaks of amino acids were very distorted. How I can improve this? [top]
When using an analytical column it is recommended to inject 1 uL to 10 uL of the biological sample which is prepared either in DI water or 50% solvent A/50% solvent B of the mobile phase. When a 100 uL injection is performed the column is overloaded.
30. I am using a Cogent Diamond Hydride column and I prepared my solvents for positive ionization LC-MS: Solvent A: 50% methanol/50% DI water + 0.1% formic acid and Solvent B: 90% acetonitrile/10% DI water/0.1% formic acid. I started the gradient from 100% Solvent A and over 10 minutes went to 50% A. Polar metabolites were not retained. What did I do wrong? [top]
ANP and HILIC gradients need to start between 100% - 90% of Solvent B (to retain polar compound) and then reduce the percent of Solvent B even down to 0% solvent B (when 50% methanol/50% water is used as Solvent A). The advantage of Cogent Type-C columns is that the equilibration between gradient runs is very fast (5 minutes) compared to HILIC columns.
31. I am using a Cogent Diamond Hydridecolumn with LC-MS and solvents containing 10 mM ammonium acetate. I noticed that my source was getting white film on it. At the same time the background noise was very low. Is my column losing the packed material? [top]
Absolutely not. Cogent Type-C columns have a very low background even when used with LC-MS. The white film comes from ammonium acetate build up when used in the mobile phase.
32. I was analyzing maleic and fumaric acids in biological extract using Cogent DH column. After 10 perfect injections (%RSD of retention times around 0.5), I noticed that the peaks intensity diminished about 10 times, however the peaks still have the same retention time. What happened to the column? [top]
Absolutely nothing happened to the column. Your mass spectrometer needs some attention. Please clean you ion source, check the nebulizer.
33. A Cogent Diamond Hydride column was used for analyzing AMP, ADP, ATP, UDP and GPT in biological extracts. After 2 days of perfect chromatograms my peaks started to deteriorate. What can I do to get back initial results? [top]
Phosphate containing metabolites are very sensitive to the presence of sodium in your LC-MS system.
After few days there is enough sodium in your system (leaching from the glass bottles) that it is sufficient to affect peak shape of this class of metabolites. There are two solutions to this problem. You can:
A: prepare fresh mobile phase daily or
B: switch to Teflon bottles.
We recommend point B for long term solution. If you happen to have J.T. Baker Nitric Acid – ULTREX made in Canada, it comes in perfect Teflon bottles. To prepare these bottles for use with LC-MS, rinse them with DI water (3 times), fill with methanol/DI water 1:1 and keep this solution overnight in the bottle. After this the bottles are ready for LC-MS
|