MicroSolv manufactures HPLC columns, autosampler vials for HPLC & GC, syringe filters, fittings and accessories and small bore tubing. CE, TLC, Flash Chromatography and OPLC are also available. Separation sciences leaders.
Log In | Register Now

       Help

 


















 


 



a1-acid-glycoprotein
      From Sheep

charts
PDF
Click here to view
printable Application Sheet

Notes: Glycoproteins are macromolecules composed of a polypeptide covalently bonded to a carbohydrate molecule. The sugar can make up anywhere from less than 1 percent to 80 percent of the total mass of a glycoprotein.
Method Conditions



Column Cogent Bidentate C8 300™, 5µm, 300A
Catalog No. 40008-75P-3M
Dimensions 4.6 x 75 mm
Solvents A: DI water + 0.1% trifluoroacetic acid (TFA)
B: acetonitrile + 0.1% TFA
Gradient
Time    %B
0.0 15
20.0 50
25.0 50
25.1 15
Post time 5 min
Flow rate 0.5 mL/min.
Sample Peak a1-acid-glycoprotein, sheep, 104H9320 0.5 mg/mL in DI water
Detection UV 214 nm
Injection 1 microL


Discussion


Glycoproteins have very diverse functions in the living organism. They are used as protective agents, lubricants and they are constituents of the cell wall. This diversity is a direct function of their structure. A simple, linear Reverse Phase gradient used in this note shows the retention of sheep glycoprotein. The resulting chromatograms are very reproducible and the equilibration time between gradient runs is only 5 minutes.