Forced Degradation of Atorvastatin
      Separation of API from its main degradation product

charts
PDF
Click here to view
printable Application Sheet


Note: Atorvastatin is a competitive inhibitor of 3-hydroxy-3-methlyglutaryl coenzyme A (HMG-CoA) reductase, which catalyzes the rate-limiting step in cholesterol biosynthesis. As such, atorvastatin is used to reduce plasma levels of low-density lipoprotein (LDL) cholesterol, which are known to contribute to the development of atherosclerosis. Atorvastatin is currently marketed by Pfizer under the trade name Lipitor, with the U.S. patent set to expire in November 2011.
Method Conditions



Column Cogent Bidentate C18™, 4µm, 100A
Catalog No. 40018-15P
Dimensions 4.6 x 150 mm
Solvents A: DI water / 10 mM ammonium acetate
B: 90% acetonitrile/10% DI water / 10 mM ammonium acetate.
Both solutions were vacuum filtered through a 0.45 µm nylon filter (MicroSolv Technology Corp. Eatontown, NJ, USA).
Gradient
time (min.) %B time (min.) %B
0 40 6 40
5 100    
Flow rate 0.8 mL/min
Injection Volume 2 µL
Sample A. Atorvastatin tablet extract
B. Degraded atorvastatin tablet extract
40 mg strength tablet ground, dispensed in 50 mL B in 100 mL volumetric flask.Vortexed 5 min, sonicated 5 min, then diluted to mark with A. Filtered through a 0.45 µm nylon membrane (MicroSolv Technology Corp. Eatontown, NJ, USA). Diluted 4x with A) 50:50 A:B or B) 1.0 N HCl in 50:50 A:B.
Peaks 1. Atorvastatin, 2. Degradant
Detection UV 248 nm

Discussion

Atorvastatin is separated from its main degradation product using a Bidentate C18 column and a simple linear reverse phase gradient. The degradant is formed under acidic conditions and therefore its separation is important for stability indicating methods for pharmaceutical formulations of atorvastatin. For methods requiring non-degrading conditions, it is important to choose diluent and mobile phase conditions with this pH sensitivity in mind. As such, ammonium acetate was chosen for the diluent and mobile phase additives in order to avoid any degradation from these sources. Data obtained under non-degrading conditions is shown in Figure A. Figure B shows an overlay of five consecutive runs of the degraded extract. The repeatability of the method is demonstrated by the low atorvastatin retention time %RSD of 0.10.