MicroSolv manufactures HPLC columns, autosampler vials for HPLC & GC, syringe filters, fittings and accessories and small bore tubing. CE, TLC, Flash Chromatography and OPLC are also available. Separation sciences leaders.
Log In | Register Now

       Help

 


















 


 



GlycoProtein Standard
From Bovine Serum

charts
PDF
Click here to view
printable Application Sheet

Method Conditions



Column Cogent Bidentate C8 300™, 5µm, 300A
Catalog No. 40008-75P-3M
Dimensions 4.6 x 75 mm
Solvents A: DI water + 0.1% trifluoroacetic acid (TFA)
B: acetonitrile + 0.1% TFA
Gradient
Time (min.)    %B
0.0 15
20.0 50
25.0 50
25.1 15
Post time 5 min
Flow rate 0.5 mL/min.
Sample Prep 1. a1-acid-glycoprotein, bovine serum, 065K705 0.05% w/v in DI water.
Detection UV 214 nm.
Injection 1 microL


Discussion


High performance liquid chromatography (HPLC) is the method of choice for the analysis of proteins and glycoproteins. The figure shows a chromatogram analyzing a solution of bovine glycoprotein standard (Sigma). The method can be used for measuring relative and absolute contents of glycosylated proteins in plasma or urine samples. Quantitation of the individual glycoproteins is straightforward. The Cogent Bidentate C8 300 column is an excellent choice for gradient analysis due to short equilibration time between runs and repeatability (precison) of the results.






[TOP]


Privacy Statement
MicroSolv Technology Corporation on LinkedIn Site developed by Net Ink Designs.