CapLC Application Sheet
Separate of Hormones with Cogent Bidentate C18™ Capillary LC Column in Reverse Phase

Shorter Analysis Time

charts

PDF
Click here to view
printable Application Sheet
Method Conditions



Column Cogent Bidentate C18 capillary LC Column, 4µm, 100A
Catalog No. 40018-A-50
Dimensions 50 µm i.d. x 40 cm packed
(50 cm total length)
Mobile Phase 70:30 acetonitrile/DI water + 0.1% formic acid
Flow rate 100 nL/minute
Injection Volume  10 nL
Sample    1. Prednisolone
   2. Corticosterone
   3. Norgestel
   4. Progesterone
      100 mg/mL of each in acetonitrile + DI water
Detection UV 254 nm
Instrument Trisep™ - 2100 EPLC system, Unimicro Tech. Inc.


Discussion


The chromatogram presented was done in Capillary LC and shows the separation of mixture of four steroids. The peak shape of each compound is excellent (As < 1.1 for all solutes) using an isocratic mobile phase. The Signal to Noise ratio is excellent, due to a high order digital filter used in the detector. Small peaks of impurities and degradation products of hormones tested are visible. The instrument is a valuable addition to any analytical laboratory.