HPLC Application Sheet
Normal Phase Made Easier: Drying Solvents in the Mobile Phase not Necessary
Method Conditions
| Column |
Cogent Bidentate C18, 4µm, 100A |
| Catalog No. |
40018-75P |
| Dimensions |
4.6 x 75 mm |
| Mobile Phase |
95:5 Hexane/Ethyl Acetate
|
| Flow rate |
1.0 mL/minute |
| Injection Volume |
1 µL |
| Samples |
1 mg/mL of proprietary compounts were dissolved in the mobile phase |
| Peaks |
- Phenolic compound with an aldehyde
- Parent phenolic compound
- Phenolic compound with a ketone
- Phenolic compound with an acid group
|
| Detection |
Photo Diode Array (UV) |
Discussion
This Cogent Bidentate C18 column was used to quickly separate four proprietary phenolic compounds
(precursors for a catalyst) under Normal Phase conditions. The separation shown is extremely reproducible
(%RSD 0.2) and the solvents do not need to be dried before use as with other normal phase columns, saving
time and lab resources.
Notes:
- This method can also be used with Atmospheric Pressure Chemical Ionization in the positive or negative
mode: APCI+, APCI-
- The column used for this application was previously used under Reversed Phase condition. Switching the
column to Normal Phase mode required flushing 100% methanol for 10 minutes followed by 100% Methylene
Chloride for 10 minutes. This process can be repeated over and over without damage to the column and can go
back and forth between modes.
|