Purine Nucleotides
Fast Separation of XMP and GTP

Click here to view printable Application Sheet
Notes:
Among the most important purine nucleotides, XMP and GTP are often analyzed from the
lymphocytes of healthy people and HIV-1 seropositive patients at different stages of
the disease (ARC-AIDS). Several differences in metabolism of purine nucleotides in
the lymphocytes of the AIDS patients, were observed [2]. XMP does not normally appear
in free nucleotide cell extracts, however it is the product of the important cell
differentiation enzyme IMP dehydrogenase in ribavirin therapies.
[1]. “Aqueous normal phase retention of nucleotides on silica hydridebased columns:
Method development strategies for analytes relevant in clinical analysis”, M.T.
Matyska, J.J. Pesek, J. Duley, M. Zamzami, S.M. Fisher, J. Sep. Sci. (2010/, 33,
930-938.
[2]. “Analysis of purine nucleotides in lymphocytes from healthy subjects and AIDS
patients”, A Tabucchi, F Carlucci, E Ramazzotti, MC Re, F Marinello, M Rubino, R
Pagani, Biomedecine & Pharmacotherapy, Volume 46, Issue 1, 1992, Pages 25-29.
|
|
Method Conditions
| Column |
Cogent Diamond Hydride, 4µm, 100A |
| Catalog No. |
70000-15P-2 |
| Dimensions |
2.1 x 150 mm |
| Solvents |
A: DI water + 0.1% ammonium formate
B: 90% acetonitrile/10% DI water/0.1% ammonium formate |
| Gradient |
| Time (min) | %B |
| 0.0 | 95.0 |
| 0.5 | 95.0 |
| 10.0 | 75.0 |
| 15.0 | 30.0 |
| 20.0 | 30.0 |
| 20.1 | 95.0 |
|
| Post time |
5 min. |
| Temperature |
25°C |
| Flow rate |
0.4 mL/min. |
| Injection |
1µL |
| Samples |
- XMP: Xanthosine -5'-monophosphate
- GTP: Guanosine -5'-triphosphate
|
| Detection |
UV diode array
using chromatograms at 254 nm and 280 nm, extracted from the UV spectra
scanned from 230 to 310 nm for all runs. |
Discussion
This fast, simple and easy to use method achieves the separation of nucleotides XMP from GTP.
The analysis of these polar compounds is achieved at high concentration of an organic solvent
as part of the mobile phase (usually acetonitrile, but acetone may be used as well when MS
detection is used) which provides increased sensitivity. Retention times were very reproducible
with %RSD approximately 0.4, even when red blood cells extracts were injected in between the
standard samples. Using a mass sprectometer this method can be specific and sensitive.
|