HPLC Application Sheet
Separation of Steroids by LC-MS

charts

PDF
Click here to view
printable Application Sheet

Method Conditions



Column Cogent Qx2™ Column for LC-MS
UDC-Cholesterol, 4 µm, 100A
Catalog No. 69069-Q20
Dimensions 2.1 x 20 mm
Mobile Phase  35:65 methanol/DI water + 0.5% formic acid
Flow rate 0.3 mL/minute
Injection Volume 1 µL
Samples
  1. Adrenosterone (m/z 301)
  2. Corticosterone (m/z 347)
  3. Estrone (m/z 253)
  4. 4-Adrostene-3, 17-dione (m/z 287)
    (10 µg/mL of each dissolved in methanol)
Detection Atmospheric Pressure Chemical Ionization in positive mode: APCI+ Single Ion Monitoring


Discussion


Separation of four steriods in the reversed phase mode using the UDC-Cholesterol Qx2™ column is easily acheived. Retention order is determined by hydrophobic interactions as well as the shape selectivity of the cholesterol stationary phase. Note that the mobile phase used was isocratic.