Instrument: |
|
Agilent 1100 LC System #9 |
Agilent 1100 LC System #2 |
|
Performance Characteristic: |
Test Conditions: |
Performance: |
Performance: |
|
System Extra-Column Dispersion: |
Rs test mix extrapolated N |
5 uL |
3 uL |
|
Gradient Dwell Volume: |
Dwell Volume: |
1.20 mL |
1.07 mL |
|
|
Delivery Accuracy: |
All circuits <5% of nominal |
All circuits <5% of nominal |
|
% Injector Carryover: |
No Wash Vial: |
0% |
0% |
|
With Wash Vial: |
0% |
0% |
|
Injection Precision: |
10 replicate inj. |
0.07% RSD |
0.07% RSD |
|
Ret. Time Precision: |
10 replicate inj. |
0.21% RSD |
0.06% RSD |
|
System Sensitivity: |
caffeine @ 273nm normalized for column efficiency
peak to peak noise @ 273 nm |
Sens. = 363.9 mAU/µg
LOD = 0.032 µg/mL
Noise = 0.04 mAU |
Sens. = 400.6 mAU/µg
LOD = 0.058 µg/mL
Noise = 0.077 mAU |
|
Pump Flow Rate Accuracy: |
0.5 mL/min |
100.3% |
100.0% |
|
1.0 mL/min |
100.3% |
100.3% |
|
5.0 mL/min |
99.7% |
99.7% |
|
Pump Max Press: |
BPR, leak check |
Qual: to 5,000 psi, 345 bar |
Qual: to 5,000 psi, 345 bar |
|
Column Oven: |
oven air temp <5°C of normal |
Qual: 20-50°C, <5°C of normal |
Qual: 20-50°C, <5°C of normal |