Improve Peak Shapes in LC-MS: Add 10 µM EDTA to the Mobile Phase
Poor peak shapes—especially for polyprotic acids and nucleotides—can often be traced back to metal contamination. These metals may leach from:
- Glass solvent bottles
- Autosampler vials
- Internal components of the LC-MS system
The extent of this issue can vary significantly between instruments, so it's important to identify the specific source of peak distortion in your setup.
Suggested Solution
To mitigate metal-related peak tailing or distortion, consider adding 10 µM EDTA to your mobile phase when using Cogent TYPE-C HPLC columns. These chelating agents can bind trace metals, improving peak symmetry and reproducibility.
Note: Always verify compatibility with your method,column and instrument before modifying mobile phase composition.