Ticagrelor Analyzed with HPLC - AppNote
September 10, 2021
/
/
/
A Reproducible Method for Detection of this Blood Clot Preventative
Click HERE for Column Ordering Information.

A rapid, sensitive, and Reproducible Method has been developed for analysis of Ticagrelor. The data below, (overlay of 5 chromatograms ) illustrates how the compound can be adequately Retained and detected using a simple Gradient in Reversed Phase HPLC.  The Method demonstrates good Peak Shape and run-to-run Precision with RSD values less than 0.3%.

A Phenyl ring in the Column Stationary Phase provides beneficial π-π Interaction with the Analyte making possible the use of a very simple, Mass Spec friendly Mobile Phase with Formic Acid as an additive.

5 Injections of Ticagrelor Overlay


Method Conditions
Column: Cogent Phenyl Hydride™, 4 μm, 100 Å
Catalog No.: 69020-10P-2
Dimensions: 2.1 x 100 mm

Mobile Phase
:
  • A: DI Water with 0.1% Formic Acid (v/v)
  • B: Acetonitrile with 0.1% Formic Acid (v/v)
Gradient:
Time (minutes) %B
0 40
1 40
5 85
6 85
7 40
10 40

Injection vol.
: 1 μL
Flow rate: 0.2 mL / minute
Detection: UV @ 254 nm

Sample Preparation
: 2.0 mg / mL Ticagrelor standard solution in Methanol.

Notes : Ticagrelor is an oral prescription indicated for treatment of acute ischemic stroke, coronary artery disease and acute coronary syndrome.
 