What Are the Thread Specifications for The ARE Brand PEEK HPLC Column Hardware - FAQ
December 20, 2012
/
/
/
Date: 20-DECEMBER-2012   Last Updated: 28-NOVEMBER-2025

The PEEK column hardware has two specifications for threading:

 
1. The female outlet threading found on the outside of the end fittings, where the end plugs or extra-column tubing ferrules connect. This threading is standard 10-32 for all items regardless of internal diameter (ID). This accommodates capillary tubing that is 1/16" outer diameter (OD).

2. The threading that connects the column tube to the end fittings. This varies depending on ID and is 1/4 - 28 for 2mm ID columns,  1/2 - 20 for 4.6mm, and 5/8 - 18 for 7.5mm.