Shorter Column Supplied in the HSQ Kit Benefits - Tech Information
April 9, 2014
/
/

Date: 9-APRIL-2014   Last Updated: 1-JULY-2025

The HSQ Kit™ comes with a 4.6 x 75mm column that is used in the qualification procedure.

The older version of the the HSQ Kit™ (called the PQ Kit™) comes with a 4.6 x 150mm column instead.

This difference will not affect the qualification data results and the columns can be used interchangeably. The only difference for the end-user is that the HPLC runs take much less time due to the shorter column length when using the HSQ Kit™. Hence, this anadvantage over the older PQ Kit.

Click HERE for HSQ & PQ kits ordering information and pictures.

Related Articles

  1. Differences in the HPLC Qualification Kits - Tech Information
  2. Linearity Solutions in the Chemical Solutions HPLC Qualification Kits - Tech Information
  3. SOP Templates are not Included with Chemical Solutions Qualification Kits - Tech Information