Avoid Spurious Peaks from Nylon Syringe Filters with DI Water Using Electrochemical Detection - Tips & Suggestions
March 19, 2015
/
/
/
Date: 19-March-2015   Last Updated: 3-NOVEMBER-2025

When using electrochemical detection, spurious peaks in your chromatogram may be observed if a nylon syringe filter is used with a DI water solvent. The anionic peaks are most likely nitrate extractable and therefore we recommend using polyester sulfone PES syringe filters instead. This membrane type is suited to situations when you want to avoid nitrate extractables. Click HERE for PES syringe filter ordering information and pictures.

If anions are not a concern, nylon filters are adequate for aqueous solvents with electrochemical detection.

Click HERE for AQ brand syringe filter ordering information and pictures.