Precision in Analytical Analysis Definition Made Easy - PRIMER
April 14, 2020
/
/
/
Date: 14-APRIL-2020   Last Updated: 26-OCTOBER-2025

The precision of a chromatographic method is determined by using different levels of analyte concentrations (the same concentration levels taken in the accuracy study), prepared from independent stock solutions and analyzed (n=10). Inter-day, intra-day and inter-instrument variation should be studied to determine the intermediate precision of the proposed analytical methods.

Different levels of analyte concentrations in triplicates should be prepared three different times in a day and studied for intra-day variation. The same procedure should be followed for three different days in order to study the inter-day variation (n=10). One set of different levels of the concentrations should be reanalyzed using another HPLC system, by proposed methods to study inter-instrument variation (n=10). The percent relative standard deviation (RSD %) of the predicted concentrations from the regression equation is taken as precision.

 