Adenosine USP Assay Using RP Hyper Similars C18 Columns - AppNote
March 21, 2025
/
/
/
/

Date: 31-MARCH-2025   Last Updated: 1-SEPTEMBER-2026

Alternate L1 Column used for a USP Assay 

This application note illustrates the USP method for Adenosine using the Cogent RP Similars C18 ODS column. The chromatograms demonstrate that the column meets the USP system suitability requirements, including tailing factor, resolution, and relative standard deviation. This makes it an excellent alternative for a USP Adenosine assay.

Peak 1.   Adenosine

Method Conditions:
Column: Cogent RP Hyper Similars C18™ 5 um, 100A
Catalog No.: 62518-25P
Dimensions: 4.6 mm x 250 mm

Sample Preparation: --  0.2 mg/mL of Adenosine in Mobile phase
System suitability solution: 4 μg / mL each of Adenine and inosine in Mobile phase (peak order respectively.) 

Resolution: ≥ 1.5 between adenine and inosine,
Tailing factor: ≤ 0.99 Standard solution
Relative standard deviation: 0.5% Standard solution


Note 1: For the full method conditions and details, consult the official United States Pharmacopeia–National Formulary (USP–NF.)


© Copyright 2026. MICROSOLV. All Rights Reserved. Website & Hosting by BlueTone Media