Separation of API from Matrix Peaks from an Over The Counter Tablet
This Method for Analysis of a Bisacodyl Tablet formulation Separates the API from several small matrix Peaks. These Peaks must be Resolved from the main Peak in order to ensure accurate Quantitation. The data below illustrates the separation potential of the this Method for Assay or Impurity Methods. Ammonium Acetate was used in the Mobile Phase and extraction Solvent to avoid any acid-catalyzed hydrolysis of the Bisacodyl ester groups.
Runs from two Column lots are shown demonstrating consistency and Robustness of this Method.
Method Conditions
Column: Cogent Bidentate C18™, 4 μm, 100 Å
Catalog No.: 40018-75P
Dimensions: 4.6 x 75 mm
Mobile Phase:
--
A: DI Water with 10 mM Ammonium Acetate
--
B: 95:5 Acetonitrile / Solvent A (v/v)
Gradient:
| Time (minutes) | %B |
|---|---|
| 0 | 20 |
| 1 | 20 |
| 6 | 60 |
| 8 | 60 |
| 9 | 20 |
Injection vol.: 2 μL
Flow rate: 1.0 mL/minute
Detection: UV @ 254 nm
Sample Preparation: 5 mg strength Bisacodyl Tablet was ground and added to a 10 mL volumetric flask containing a portion of Solvent B. It was then sonicated 10 minutes and diluted to mark. After mixing, a portion was filtered with a 0.45 μm Nylon Syringe Filter.
t 0 : 0.9 minutes
Note: Bisacodyl is a stimulant laxative used in many common over-the-counter formulations. Brand names include Dulcogen®, Alophen®, and Dulcolax®.