Can I use the Chemical Solutions HSQ Kit with the longer column from the original PQ Kit - FAQ
May 15, 2014
/
/
/
/

Date: 15-MAY-2014   Last Updated: 5-JULY-2025

🔄  Yes, you can — with a few simple adjustments.

If you're required to use the older 150mm PQ™ column instead of the standard 75mm HSQ™ column, the HSQ Kit can still be used effectively by modifying a few parameters in the included Excel spreadsheet.


🛠️ Recommended Adjustments:

To ensure accurate results with the longer column, update the following settings:

Parameter

Original HSQ Setting

Adjusted for 150mm Column

Flow Rate

2.0 mL/min

1.0 mL/min

Retention Times

Standard

Multiply by 4×

Injection Volume

8 µL

Increase to 12 µL

Gradient Test Steps

5-minute steps

Extend to 10-minute steps (for systems with large dwell volumes)

These changes compensate for the longer column length and lower flow rate, ensuring your qualification results remain valid and reliable.


⚠️ Important Notes:

  • All other test parameters and results remain unchanged between the HSQ and PQ kits.
  • Using the 75mm HSQ column is recommended when possible, as it reduces run time by approximately 75%, saving time and resources.

📦 Click HERE for HSQ & PQ Kit ordering information and product images

Need help modifying the spreadsheet or optimizing your setup? Our technical team is here to assist!

 

Related Articles

  1. Can Chemical Solutions Qualification Kits Be Used with Any HPLC Instrument - FAQ
  2. How does the shorter column supplied in the HSQ kit affect the data compared to the longer column from the PQ Kit - FAQ
  3. What is the difference between the HSQ Kit and older PQ Kit - FAQ

© Copyright 2025. MICROSOLV. All Rights Reserved. Website & Hosting by BlueTone Media