Cleaning the PQ or HSQ Kit HPLC Column - Tech Information
January 21, 2016
/

Date: 21-JANUARY-2016   Last Updated: 5-JULY-2025

🧪 Routine Use:

  • Before starting your qualification sequence, a 10–15 minute equilibration is typically sufficient.
  • No additional cleaning is usually necessary if the column is used as directed.

🧼 If Cleaning Is Ever Required:

  • Flush the column with a 50:50 mixture of acetonitrile and water for 15–30 minutes.
  • This gentle rinse is effective for removing any incidental buildup without compromising column integrity.

 


NOTE:

  • Maintaining the column in clean condition ensures consistent and reliable qualification results. Do not use these columns for any other purpose.
  • The HPLC columns included in the Chemical Solutions™ PQ or HSQ Qualification Kits are intended exclusively for instrument performance qualification testing, not routine sample analysis. As such, it should remain uncontaminated under normal use.