Column Protection Suggestion for a 1mm Cogent 2.o Column - Tech Information
April 11, 2018
/
/
/
/
Date: 11-APRIL-2018   Last Updated: 11-FEBRUARY-2026

Which Guard Column Should You Use with a 1 mm Cogent 2.o Column?

Users of 1 mm internal‑diameter Cogent 2.o™ columns—which feature 2.2 µm near‑UHPLC silica hydride particles—often look for a matching guard column to protect their analytical investment.

However, no guard column exists with the exact same internal dimensions required for proper flow and performance. Because of this, a guard column is not recommended for these ultrahigh‑efficiency narrow‑bore columns. 


Why No Guard Column Is Offered for the 1 mm Cogent 2.o Format

Guard columns must match the analytical column’s:

  • Internal diameter (ID)
  • Particle size
  • Packing characteristics
  • Flow profile

For the 1 mm ID / 2.2 µm geometry, no guard can maintain identical backpressure and flow distribution. Any mismatch would compromise:

  • Peak shape
  • Reproducibility
  • Efficiency
  • Sensitivity (especially in LC–MS)

Thus, no guard should be used with this specific column format. 


Recommended Alternative: Cogent Column Filters

Instead of a guard, we recommend protecting 1 mm columns using Cogent Column Filters, which are:

  • Disposable
  • Inline
  • Non‑retentive (they do not change chromatography)
  • Designed solely for particulate protection

These filters safeguard the column from:

  • Undissolved sample particulates
  • Mobile‑phase debris
  • Particles from pump seal wear
  • Accidental contamination during routine operation

Recommended Part Number

For 1 mm ID columns with 2.2 µm particles, we recommend: Cogent Column Filter – Item No. 49650‑005

This filter setup effectively replaces the need for a guard column without disrupting performance.


Why Filters Work Better Than Guards for Ultra‑Narrow Columns

1. Prevent Backpressure Issues

Guard columns add extra length, extra packing, and extra restriction—often more than a 1 mm ID system can tolerate while maintaining method integrity.

2. Protect Without Affecting Chromatography

Filters do not retain analytes; they simply trap particulates upstream.
we emphasize that chromatography remains unchanged, making them a safer alternative.

3. Lower Cost, Higher Reliability

Filters are disposable and inexpensive compared to guard cartridges. More importantly, they prevent irreversible damage to the narrow inlet frits of 1 mm columns, which are extremely sensitive to clogging.


For the 1mm ID column with 2um particle size we recommend item number 49650-005
Click HERE for ordering information and a picture of the filter. 

 
Cogent HPLC Column Logo

Related Articles

  1. Cogent Column Filters Do Not Need Special Fittings to use with HPLC column - Tech Infomation
  2. Cogent Column Filters Should Not Replace HPLC Guard Columns - Tech Information
  3. Which solvents should not be used with the Cogent Column Filters - FAQ

© Copyright 2026. MICROSOLV. All Rights Reserved. Website & Hosting by BlueTone Media