Range the UV-vis Calibration Kit Operates - Tech Information
July 29, 2020
/

Date: 29-JULY-2020   Last Updated: 28-JUNE-2025

The MICROSOLV UV-Vis Calibration Kit™ is designed to support calibration in the UV (200–400 nm) and Visible (400–780 nm) spectral regions. It includes:

  • Potassium dichromate for absorbance accuracy
  • Holmium oxide for wavelength accuracy
  • Potassium iodide for stray light assessment

These components are suitable for compliance with USP <857>, but only within the wavelength ranges covered by the included solutions.

This kit does not support calibration in the Near-Infrared (NIR) range, which typically spans 780–3300 nm, depending on the spectrophotometer. NIR calibration requires different standards and procedures not included in this kit.

For reference, the chart below outlines the acceptable criteria and wavelength ranges defined in USP <857>. The highlighted section indicates the specific regions addressed by our calibration kit.\

 

Related Articles

  1. L3 Linearity Solution Useful for UV Wavelength Accuracy - Tech Information
  2. Qualification with Detection at 200 nm using UV-vis Calibration Kit - Tech Information
  3. USP-NF Chapter Revisions for 857 and 1058 Kit Compliance - Tech Information

© Copyright 2026. MICROSOLV. All Rights Reserved. Website & Hosting by BlueTone Media