Reducing Peak Broadening with Direct Adaptive Di-Ad HPLC Column Connectors - Tech Information
February 28, 2022
/

28-FEBRUARY-2022   Last Updated: 4-SEPTEMBER-2026

How to reduce peak broadening due to poor column connections

The spring-loaded "Direct Adaptive" Di-Ad™ HPLC column connectors were designed for speed and convenience when changing between different column brands on your instrument. 

In the example below, we demonstrate how this error-free process using DiAd fittings can also reduce peak broadening from common poor connections.
 
 
Trace Fitting Used Height (mAU) Plates (N)
Black PEEK 31.33 6306
Red Poorly Inserted PEEK 30.36 6023
Blue Di-Ad™ 35.37      7160
 


Method Conditions:
Direct Adaptive HPLC column connector: Double end fitting

Catalog No.: 49910-10-DD
Dimensions: 0.010" ID x 1/16th" OD, 100 mm long

Column: Cogent Bidentate C8™, 4 μm, 100 Å.
Catalog No.: 40008-10P
Dimensions: 4.6 mm x 100 mm

Mobile Phase: 40% acetonitrile / 60% DI water 0.1 % formic acid
Injection vol.: 1 µL
Flow rate: 1.0 mL / minute

Detection: UV 254 nm
Sample Preparation: 0.1 mg / mL phenol in mobile phase.



Notes: Column efficiency: theoretical plates (N) 1/2 Height    N = 5.54((tR/W)2)
tR = Retention time of peak
W = Width of peak measured at 1/2 height


 