Fructose and Glucose with HPLC & ELSD - AppNote
December 13, 2016
/
/
/
/
/
Date:  10-OCTOBER-2025     Last Updated: 10-OCTOBER-2025


The Cogent Amide Column, combined with a TEA-based mobile phase, offers excellent separation performance, and the method is well-suited for routine sugar analysis in beverages.


PEAKS:

1. D-Fructose
2. D-Glucose

Method Conditions
Column: Cogent Amide™, 4μm, 100Å
Catalog No.: 40036-10D
Dimensions: 4.6 x 100 mm
Mobile Phase:   10% DI Water / 90% Acetonitrile (v/v),
Injection vol.: 3μL
Flow rate: 1.0 mL/minute
Detection: ELSD (Gain 8, Temp 50C, Nitrogen: 3.5lb)
Sample: D-Fructose and D-Glucose reference standards (1 mg/mL) in diluent of 50% Acetonitrile / 50% DI Water / 0.1% TEA (v/v)

 




 

Related Articles

  1. Fructose and Glucose in Coca-Cola Analyzed with HPLC & RI - AppNote
  2. Galactose and Sucrose - AppNote
  3. Why am I seeing multiple peaks in HPLC runs of fructose - Tips & Suggestions

© Copyright 2025. MICROSOLV. All Rights Reserved. Website & Hosting by BlueTone Media