Fructose and Glucose with HPLC with ELSD - AppNote
December 13, 2016
/
/
/
Date:  10-OCTOBER-2025     Last Updated: 10-OCTOBER-2025

The Cogent Amide Column, combined with a TEA-based mobile phase, offers excellent separation performance, and the method is well-suited for routine sugar analysis in beverages.

 

PEAKS:
1. D-Fructose,   2. D-Glucose

Method Conditions
Column: Cogent Amide™, 4 μm, 100 Å
Catalog No.: 40036-10D
Dimensions: 4.6 x 100 mm

Mobile Phase:   10% DI Water / 90% Acetonitrile (v/v),
Injection vol.: 3 μL
Flow rate: 1.0 mL / minute
Detection: ELSD (Gain 8, Temp 50C, Nitrogen: 3.5lb)

Sample: D-Fructose and D-Glucose reference standards (1 mg / mL) in diluent of 50% Acetonitrile / 50% DI Water / 0.1% TEA (v/v)


 

Related Articles

  1. Fructose and Glucose in a Cola Analyzed with HPLC and RI - AppNote
  2. Galactose and Sucrose Refractive Index Method - AppNote
  3. Multiple peaks in HPLC runs of fructose - Tips and Suggestions