Melamine Trace Levels Analyzed with LC-MS - AppNote
May 15, 2012
/
/
/
/
Date: 15-MAY-2012 Last Updated: 22-AUGUST-2026
Detect Trace Levels of MelamineThe unique chemical properties of Melamine (base functionalities) present challenges in obtaining Retention and adequate Sensitivity, when using standard HPLC Columns. In this Method, Melamine is Analyzed under HILIC conditions with an increased Sensitivity making it possible for detection at Trace Levels.
Peak: Melamine m/z 127 (M+H)+
Method ConditionsColumn: Cogent Diamond Hydride™, 4 μm, 10 Å
Catalog No.: 70000-15P-2
Dimensions: 2.1 x 150 mm
Mobile Phase:
- I: 60% acetonitrile / 40% DI water / 0.1% formic acid
- II: 70% acetonitrile / 30% DI water / 0.1% formic acid
Detection: ESI – pos – Agilent 6210 MSD TOF Mass Spectrometer
Sample Preparation: 166.7 µg / mL of compound in 50% Acetonitrile / DI Water / 0.1% Acetic Acid. Run 1:1000 dilution in 80% Acetonitrile / DI Water / 0.1% Acetic Acid
t 0 : 1.44 minute
Notes: Several dogs and cats died or were seriously ill after consumption of pet foods which had been manufactured with adulterated wheat glutens. After testing of the wheat glutens it was determined that they contained significant levels of Melamine and / or several related compounds.