Glyburide and Metformin Analyzed with HPLC - AppNote
May 23, 2012
/
/
/
/
/
/
Separation of Highly Polar and Non-Polar Compounds in one Isocratic run

In this Method, the polar compound, Metformin, and the nonpolar compound Glyburide, can be retained on the same Stationary Phase (see A & B). Depending on the Mobile Phase composition either Metformin or Glyburide can be retained longer.

 

 


Method Conditions
Column: Cogent Bidentate C18™, 4µm, 100Å
Catalog No.: 40018-75P
Dimensions: 4.6 x 75mm
Mobile Phase:
-- A: 15% DI Water / 85% Acetonitrile / 0.5% Formic Acid
-- B: 50% DI Water / 50% Acetonitrile / 0.5% Formic Acid
Injection vol.: 1µL
Flow rate: 0.5mL / minute
Detection: UV @ 254nm
Sample Preparation: Stock Solution: 100µg / µL Glyburide and Metformin


Notes: Elution order was confirmed by LCMS, APCI+, with single ion monitoring Metformin (m/z 130) and Glyburide (m/z 369).

Attachment  No 08 Glyburide and Metformin Analyzed with HPLC pdf 0.2 Mb  Download File