Cyclobenzaprine Retention with an Inverse Gradient - AppNote
March 26, 2013
/
/
/

Date: 26-MARCH-2013   Last Updated: 20-AUGUST-2026

HILIC Retention of a Hydrophobic Compound 

This Application Note shows an example of a hydrophobic analyte retention in HILIC Like Gradient HPLC. Although Cyclobenzaprine has a logP of approximately 4.6–4.9, it is well retained with a Symmetrical Peak Shape.

Data from two production lots of Columns is shown to illustrate the Reproducibility of the Method.

 

Peak:
Cyclobenzaprine

Method Conditions
Column: Cogent Diamond Hydride™, 4 μm, 100Å
Catalog No.: 70000-7.5P
Dimensions: 4.6 x 75 mm

Mobile Phase:
-- A: DI Water with 0.1% Formic Acid (v/v)
-- B: Acetonitrile with 0.1% Formic Acid (v/v)

Gradient:
Time (minutes) %B
0 95
1 95
6 40
7 95
Injection vol.: 1 μL
Flow rate: 1.0 mL / minute
Detection: UV @ 224 nm

Sample Preparation: 10 mg strength Cyclobenzaprine Tablet was ground and added to a 25 mL volumetric flask. A diluent of 50:50 Solvent A / Solvent B was added and the flask was sonicated for 30 minutes. It was then diluted to mark, mixed, and filtered with a 0.45 μm Nylon Syringe Filter (MICROSOLV Tech Corp.). A portion of the filtrate was diluted 1:10 for HPLC Injections.

t
0
: 0.9 minutes

Note: Cyclobenzaprine is a muscle relaxant and used to treat Fibromyalgia. Cyclobenzaprine is structurally related to first-generation tricyclic antidepressants. It is marketed under the trade name Flexeril®.

Cyclobenzaprine Tablet pdf Download File


 

© Copyright 2026. MICROSOLV. All Rights Reserved. Website & Hosting by BlueTone Media