HPLC Method Transferred to Near UHPLC - AppNote
June 10, 2014
/
/
/
/
/
/
/
Date: 10-JUNE-2014   Last Updated: 7-FEBRUARY-2026

Use of near UHPLC column for improved results


This challenging compound has several amine groups and can be problematic to analyze by HPLC. Here a good peak shape is obtained using the Cogent Diamond Hydride 2.o Column. The efficiency is higher on the 2.2µm phase compared to a standard 4µm Column, leading to increased sensitivity. The method conditions are compatible with LCMS as well.

 

Peak:   Tizanidine HCI

Method Conditions
Columns:
  • Fig. A: Cogent Diamond Hydride 2.o™, 120Å
  • Fig. B: Cogent Diamond Hydride™, 4µm, 100Å
Catalog Nos.: Dimensions:
  • Fig. A: 2.1 x 50 mm
  • Fig. B: 2.1 x 50 mm
Mobile Phase:
  • A: DI Water / 0.1% Formic Acid (v/v)
  • B: Acetonitrile / 0.1% Formic Acid (v/v)
Gradient: 
Time (minutes) %B
0 95
1 95
6 40
7 95

Post time:
3 minutes
Injection vol.: 0.2 µL
Flow rate: 0.3 mL/minutes
Detection: UV @ 230 nm

Sample:
4 mg strength Tizanidine HCl tablet was ground and weighed in a 10 mL Volumetric Flask. A portion of 50/50 Solvent A / Solvent B diluent was added and the flask was sonicated 10 minutes. It was then diluted to mark and filtered with a 0.45 µm Nylon Syringe Filter (MICROSOLV Tech Corp.).

t 0 : 0.6 minutes

Note: Tizanidine is a centrally acting A2-Adren- Ergic agonist used to treat spasms, cramping, tightness of muscles, and related conditions. It is available under the trade name Zanaflex® as well as generic versions.


Attachment  
Tizanidine HCI Method Transfer pdf  Download File

 

Related Articles

  1. Atorvastatin HPLC Method Transfer to UHPLC - AppNote
  2. Riboflavin Pyridoxine Hydrophilic Vitamins Analyzed with LC-MS - AppNote
  3. Tizanidine HCl Tablet Analyzed by HPLC - AppNote

© Copyright 2026. MICROSOLV. All Rights Reserved. Website & Hosting by BlueTone Media