Cogent Diamond Hydride 2.o HPLC Column Retention Time Precision - AppNote
November 13, 2015
/
/
/

Date: 13-NOVEMBER-2015   Last Updated: 9-SEPTEMBER-2026

This graph is from a study investigating the day to day retention time precision with two test solutes on the Cogent Diamond Hydride 2.o™ HPLC column.   View Cogent Diamond Hydride HPLC Column Ordering Information


Peaks:
Blue Trace: Phenylalanine,     Red Trace: Phenylglycine

Many factors can affect column robustness and lifetime, so it is not possible to define specific values for those parameters.

Mobile Phase:
-- A: DI Water / 0.1% Formic Acid
-- B: Acetonitrile / 0.1% Formic Acid

Flow Rate: 0.3 mL / minute
Injection Volume: 0.2 uL
Detection: UV 254 nm


Gradient:
Time (minutes) %B
0 90
6 20
7 20
8 90
 