Tizanidine HCl Tablet Precision by HPLC - AppNote
September 1, 2016
/
/
/
Excellent Run to Run Precision

Tizanidine has several amine groups which can present problems to chromatographers. Residual silanols on the surface of ordinary Silica-based Columns can electrostatically interact with these groups, causing peak tailing.

The Cogent Amide Column, used in the Application Note is virtually free of residual silanols and therefore a symmetrical peak shape can be obtained. An overlay of five consecutive runs demonstrates the run- to-run precision when using these columns.

 

Peak:   Tizanidine

Method Conditions
Column: Cogent Amide™, 4 μm, 100 Å
Catalog No.: 40036-05P
Dimensions: 4.6 x 50 mm

Mobile Phase:
  • A: 90% DI Water / 10% Acetonitrile / 0.1% Formic Acid (v/v)
  • B: Acetonitrile / 0.1% Formic Acid (v/v)
Gradient:
Time (Minutes) %B
0 90
2 90
6 40
7 90

Flow rate:
 1.0 mL/minute
Detection: UV @ 230 nm
Injection vol.: 3 μL

Sample Preparation:  
4 mg strength Tizanidine HCl tablet was ground and weighed in a 10 mL volumetric flask. A portion of 50/50 solvent A/ solvent B diluent was added and the flask was sonicated 10 minutes. It was then diluted to mark and filtered with a 0.45 µm Nylon syringe filter (MICROSOLV Tech Corp). An aliquot was then diluted 1:10 with a diluent of 10% solvent A / 90%.

Note: Tizanidine is a centrally acting α2- adrenergic agonist used to treat spasms, cramping, tightness of muscles, and related conditions. It is available under the trade name Zanaflex® as well as generic versions.


Attachment  
No 356 Tizanidine Amide.pdf 0.5 Mb  Download File

 