How Can I Fix Poor Peak Shape for Phosphorylated Compounds in Aqueous Normal Phase ANP Methods - Tips & Suggestions
April 14, 2020
/
/
/
We have a number of strategies for fixing problems with peak shape for phosphorylated compounds:
  • Ensure the column is not overloaded. We recommend an injection volume of 1 µL or less with aqueous normal pase ANP methods.
  • Use Teflon bottles instead of borosilicate glass for your mobile phase reservoirs. Sodium from the glass can cause peak shape problems for anionic compounds.
  • Use 5–10 microM EDTA or medronic acid in the mobile phase and sample diluent. This will help chelate metals that cause problems for phosphates.
  • Try adding 0.01% ammonia to the sample diluent.
  • Try using a Cogent, metal free, coated stainless steel column.
     