Tizanidine HCl Tablet Analyzed by HPLC - AppNote
January 24, 2013
/
/
/
Separation Method for Tizanidine Compatible with LC-MS

Tizanidine has numerous amine functional groups and can be a challenge for analysis by HPLC. The USP method uses Phosphate in the Mobile Phase which is not compatible with LC-MS. This Method however uses Formic Acid as the Mobile Phase additive and produces a sharp, symmetrical peak.

The USP system suitability for the tailing factor is not more than 1.6, and the Peak obtained has a value of 1.1. Data from two Column lots is shown in the figure, demonstrating Robustness of this Method.

 

Peak:   Tizanidine HCI

Method Conditions
Column: Cogent Diamond Hydride™, 4 µm, 100 Å
Catalog No.: 70000-7.5P
Dimensions: 4.6 x 75 mm

Mobile Phase:
  • A: DI Water / 0.1% Formic Acid (v/v)
  • B: Acetonitrile / 0.1% Formic Acid (v/v)
Gradient:
Time (minutes) %B
0 95
1 95
6 40
7 95

Post Time:
3 minutes
Injection vol.: 1 µL
Flow rate: 1.0 mL / minute
Detection: UV @ 230 nm

Sample Preparation:
4 mg strength Tizanidine HCL tablet was ground and weighed in a 10 mL volumetric flask. A portion of 50:50 Solvent A / Solvent B diluent was added and the flask was sonicated 10 minutes. It was then diluted to mark and filtered with a 0.45 µm Nylon Syringe Filter (MICROSOLV Tech Corp.).

t 0 : 0.9 minutes

Note: Tizanidine is a centrally acting a2-adrenergic agonist used to treat spasms, cramping, tightness of muscles, and related conditions. It is available under the trade name Zanaflex® as well as generic versions.