How to Choose an Appropriate Guard Column When Using a Cogent Column - Tips & Suggestions
April 14, 2020
/
/
/
/
Date: 14-APR-2024   Last Updated: 14-NOVEMBER-2025

Two different guard column cartriges are available for Cogent TYPE-C Silica HPLC Columns: 2.0 and 4.0 mm internal diameter (ID). 
Both sizes are 20mm in length and are used with the same Guard Column Holder.

The 2.0mm ID guard columns are for use with the 2.1mm ID analytical columns and the 4.0mm ID guard columns should be used with the 3.0mm and 4.6mm ID analytical columns.

NOTE: Typically one would use the same stationary phase type in both the guard and analytical columns. However, in some instances it is preferred to use a Reversed Phase guard column (such as the Bidentate C18™) with an Aqueous Normal Phase  ANP column (such as the Diamond Hydride™) . It depends on which types of analytes from the matrix one expects to be most problematic in terms of carryover and adsorption on the analytical column.
 