Overview
PTFE (polytetrafluoroethylene) membranes are well‑known for their excellent chemical resistance and suitability for aggressive organic solvents. However, these same characteristics make PTFE poorly suited for aqueous filtration.
Understanding why water performs poorly through PTFE helps analysts select the correct membrane for efficient and reliable sample preparation.
Why PTFE Filters Water Slowly
PTFE is one of the most hydrophobic polymers used in filtration. Its surface strongly repels water due to:
- Extremely low surface energy
- Non‑polar chemical structure
- Inability of water molecules to wet or penetrate the membrane pores
Because water cannot readily wet the surface, the membrane tends to trap air inside its pores. This results in:
- Very slow flow rates
- High back pressure
- Difficulty initiating filtration
- In some cases, complete refusal to filter water
Even with significant pressure applied, the flow remains inefficient.
Hydrophilic Membranes Perform Better for Water‑Based Samples
For aqueous sample preparation—including HPLC, dissolution testing, and routine buffer filtration—hydrophilic membranes perform far better. Recommended alternatives include:
1. Hydrophilic Nylon
- Naturally hydrophilic
- Strong and durable
- Excellent for water and aqueous buffers
- Low extractables
Nylon is one of the most common choices for aqueous HPLC samples.
2. PES (Polyethersulfone)
- Highly hydrophilic
- Very low extractables
- Fast flow rates
- Ideal for biological or complex aqueous matrices
PES is often preferred when exceptionally clean filtration is required.
3. Cellulose Acetate (CA) or Other Hydrophilic Polymers
These may be used depending on sample chemistry, pH, and specific application requirements.
When to Use PTFE
PTFE remains the membrane of choice when filtering:
- Strong organic solvents
- Corrosive acids (in non‑aqueous systems)
- Aggressive chemical mixtures
- High‑pH or reactive compounds where membrane robustness is essential
However, for water‑based samples, PTFE should be avoided unless pre‑wetting with an alcohol is allowed—which is not recommended for most analytical workflows due to contamination risks.