The current change criteria for USP <621> (Chromatography) were significantly updated in December 2022 to allow greater flexibility in method adjustments while maintaining system suitability. Here's a summary of the key allowable changes we found using Copilot.:
✅ Allowed Adjustments Without Revalidation (if System Suitability Passes with changes)
Column Parameters
-
Column Length & Particle Size:
- You may adjust the column length and/or particle size as long as the L/dp ratio (length/particle size) remains within −25% to +50% of the original.
-
Internal Diameter:
- Can be changed even without altering particle size or length.
Mobile Phase Composition
- Minor components can be adjusted by ±30% relative, but no component should change more than ±10% absolute.
pH and Buffer
- pH: ±0.2 units (unless otherwise specified).
- Buffer concentration: ±10%.
Injection Volume
- Can be adjusted using a specific formula to maintain proportionality with other method parameters.
Column Temperature
- Isocratic methods: ±10°C
- Gradient methods: ±5°C
Gradient Methods
- Previously not allowed, gradient adjustments are now permitted, provided system suitability and selectivity are maintained.
Detector Wavelength
- No changes allowed.
🧪 System Suitability Requirements Must Still Be Met
Even with these adjustments, the method must still meet system suitability criteria such as:
- Resolution
- Tailing factor
- RSD of replicate injections
- Selectivity and elution order of impurities
- Signal to Noise remains within guidelines