In the strictest regulatory sense, no Excel-based tool can be fully 21 CFR Part 11 compliant on its own. Compliance with 21 CFR Part 11 requires a validated, closed system with features like:
- Secure user access
- Full audit trails
- Electronic signature controls
- Oversight by a qualified Quality Assurance (QA) system
🧪 What Does This Mean for the Excel Template?
The Excel template included with our Chemical Solutions Qualification Kits is not a standalone 21.11-compliant system. However, it is:
- Locked and protected to prevent accidental or unauthorized changes
- Validated to ensure accurate calculations for precision, linearity, and other qualification metrics
- Designed to work with data generated from your 21.11-compliant HPLC software
This means your original qualification data remains compliant, and the Excel tool simply processes that data in a controlled and reliable way.
✅ QA Acceptance
Our spreadsheet is designed to meet the expectations of most Quality Control (QC) and QA departments. A validation report is available upon request to support internal review and documentation requirements.
📄 Click HERE for PQ Kit™ ordering information