Nonylphenol Analyzed by HPLC - AppNote
September 18, 2013
/
/
/
Separation of Isomer Peaks using Normal Phase HPLC

Nonylphenol is a very hydrophobic compound and is suitable for analysis by Normal Phase. It is produced commercially by Acid-catalyzed Alkylation of Phenol with a mixture of nonenes. Therefore a variety of product isomers are possible, with different branching of the C9 group and position of the chain on the ring.

Normal Phase HPLC is well-suited to isomer separations and in this method, separation was observed between the main peak and two smaller peaks. Three runs are shown to illustrate the repeatability, which is often a concern with Normal Phase Methods that use ordinary Silica Columns.

 

Main Peak:   Nonylphenol

Method Conditions
Column: Cogent Silica-C™, 4 µm, 100 Å
Catalog No.: 40000-10P
Dimensions: 4.6 x 100 mm

Mobile Phase:

-- A: Ethyl Acetate
-- B: Hexane

Gradient:
Time (minutes) %B
0 100
4 100
19 90
20 100

Post Time:
3 minutes
Injection vol.: 1 µL
Flow rate: 1.0 mL / minute
Detection: UV @ 277 nm

Sample Preparation:
Nonylphenol reference standard dissolved in a Hexane diluent.

t
0 :
1.3 minutes

Note: Nonylphenols are used in synthesis as a starting material for various surfactants. They are subjected to Ethoxylation to produce Alkyl-Phenol Ethoxylates.


Attachment  
No 264 Nonylphenol HPLC Method pdf  0.3 Mb  Download File

 