🔄 Yes, you can — with a few simple adjustments.
If you're required to use the older 150mm PQ™ column instead of the standard 75mm HSQ™ column, the HSQ Kit can still be used effectively by modifying a few parameters in the included Excel spreadsheet.
🛠️ Recommended Adjustments:
To ensure accurate results with the longer column, update the following settings:
|
Parameter |
Original HSQ Setting |
Adjusted for 150mm Column |
|---|---|---|
|
Flow Rate |
2.0 mL/min |
1.0 mL/min |
|
Retention Times |
Standard |
Multiply by 4× |
|
Injection Volume |
8 µL |
Increase to 12 µL |
|
Gradient Test Steps |
5-minute steps |
Extend to 10-minute steps (for systems with large dwell volumes) |
These changes compensate for the longer column length and lower flow rate, ensuring your qualification results remain valid and reliable.
⚠️ Important Notes:
- All other test parameters and results remain unchanged between the HSQ and PQ kits.
- Using the 75mm HSQ column is recommended when possible, as it reduces run time by approximately 75%, saving time and resources.
📦 Click HERE for HSQ & PQ Kit ordering information and product images
Need help modifying the spreadsheet or optimizing your setup? Our technical team is here to assist!