Asymmetric Dimethylarginine ADMA 4um vs 2.o Diamond Hydride Columns - AppNote
May 29, 2013
/
/
/
ADMA can be retained using an ANP gradient method with the Cogent Diamond Hydride™ HPLC column.

Retention times for ADMA differed only slightly between the 4um and 2.o™ stationary phases. Efficiency however was notably higher when using the 2.o™ phase. This can be readily observed from the greater peak height using the 2.o™ column in the chromatogram overlay comparison shown below:

 


Method Conditions:

Column: Cogent Diamond Hydride
Column dimensions: 2.1 x 50mm (both)
Flow rate: 0.3mL/min
A: DI H2O + 0.1% formic acid
B: Acetonitrile + 0.1% formic acid
 
Time (minutes) %B
0 90
5 30
8 30
10 90
post time 4 min
 


For more information:  Cogent Diamond Hydride Product Page

Related Articles

  1. How does efficiency in Aqueous Normal Phase ANP compare to HILIC in HPLC - FAQ
  2. Improving peak shapes in an HPLC method in Aqueous Normal Phase ANP - Tips & Suggestions

© Copyright 2026. MICROSOLV. All Rights Reserved. Website & Hosting by BlueTone Media