Guide to Store and Condition Cogent TYPE-C HPLC Columns - Tech Information
June 22, 2013
/
/
/
/
/
/
/

Preparing and Storing Cogent TYPE-C Silica™ HPLC Columns

This guide explains best practices for initial preparation and proper storage of Cogent TYPE-C Silica™ columns to maintain performance and prevent degradation—whether for overnight or long-term storage.


Storage Solvents by Mode

  • Reversed Phase (RP) or Aqueous Normal Phase (ANP):
    Store in 100% Acetonitrile if these modes will be used next.

  • Organic Normal Phase (NP):
    Store in 100% Hexane for NP applications.


Important Note on Mode Switching

Because many organic NP solvents are immiscible with aqueous RP/ANP solvents, switching modes requires an intermediate solvent. Use methylene chloride (mutually miscible) for 15 column volumes before re-equilibrating in the next mobile phase. This simple 30-minute procedure prevents phase incompatibility issues.


Mode Conversion Procedures

  • RP → NP:

    1. Pump 100% Methanol for 15 min at 1 mL/min.
    2. Pump 100% Methylene Chloride for 15 min.
      Column is ready for NP equilibration.
  • NP → RP:

    1. Pump 100% Methylene Chloride for 15 min at 1 mL/min.
    2. Pump 100% Methanol for 15 min.
      Column is ready for RP equilibration.

Why TYPE-C Silica™ is Different:
Minimal water adsorption allows fast equilibration between organic/aqueous phases and pH changes—much quicker than Type-A or Type-B silica columns.


Critical Storage Guidelines

  • Store at room temperature—never refrigerate or heat.
  • Never let the column dry out.
  • For RP use, do not store in DI water without 0.1% formic acid.

Storage Procedures

For RP or ANP Use

  • Overnight:
    1. Flush with 8–10 column volumes of 100% Acetonitrile.
    2. Cap both ends securely with provided plugs.
  • Long-Term:
    1. Flush with 15–20 column volumes of 100% Acetonitrile.
    2. Cap ends and return to original box.
    3. Store at room temperature.

** Diamond Hydride™ columns have a special storage protocol, given click HERE.


For Organic NP Use

  • Overnight:
    1. Flush with 8–10 column volumes of 100% Hexane (or other non-polar solvent).
    2. Cap ends securely.
  • Long-Term:
    1. Flush with 15–20 column volumes of 100% Hexane.
    2. Cap ends, return to box, and store at room temperature.
    3. Label the column for NP use and note the solvent inside
 

Related Articles

  1. How to Condition Cogent RP or HPS Columns for Storage - Tips & Suggestions
  2. Store and Maintain Cogent Diamond Hydride HPLC Columns - Tech Information